Chapter 5 Error Messages
Execution Errors
118
-221
Settings conflict
Indicates that a legal program data element was parsed but could not be
executed due to the current device state.
-222
Data out of range
A numeric parameter value is outside the valid range for the command.
Example:
TRIG:DEL -3
-223
Too much data
A character string was received but could not be executed because the string
length was more than 40 characters. This error can be generated by the
CALibration:STRing
command.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for the
command. You may have used an invalid parameter choice.
Example:
DISP:STAT XYZ
(XYZ is not a valid choice).
-330
Self-test failed
The power supply’s complete self-test failed from the remote interface (
*TST?
command). In addition to this error, more specific self-test errors are also
reported.
See also ‘‘Self-Test Errors’’, starting on page 120
.
-350
Queue overflow
The error queue is full because more than 20 errors have occurred. No
additional errors are stored until you remove errors from the queue. The error
queue is cleared when power has been off, or after a
*CLS
(clear status)
command has been executed.
-410
Query INTERRUPTED
A command was received which sends data to the output buffer, but the output
buffer contained data from a previous command (the previous data is not
overwritten). The output buffer is cleared when power has been off, or after a
*RST
(reset) command has been executed.
Summary of Contents for E364 A Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Page 28: ...2 General Information...
Page 40: ...3 Front Panel Operation and Features...
Page 68: ...4 Remote Interface Reference...
Page 114: ...5 Error Messages...
Page 124: ...6 Application Programs...
Page 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Page 136: ...7 Tutorial...
Page 147: ...Chapter 7 Tutorial Remote Programming 146...
Page 148: ...8 Specifications...
Page 156: ...Appendix Service Information...
Page 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Page 193: ...Appendix Service Information Replaceable Parts 192...
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......