Chapter 1 Quick Start
Line Voltage Conversion
21
1
Line Voltage Conversion
W a r n i n g
Shock Hazard
Operating personnel must not remove power supply covers.
Component replacement and internal adjustments must be made only by
qualified service personnel.
Line voltage conversion is accomplished by adjusting two components: the line
voltage selection switch and the power-line fuse on the rear panel.
1
Remove AC line power.
2
Remove the cover (Refer to General Disassembly on page 164).
3
Set two sections of the line voltage selector switch on the PC board for the
desired line voltage (See Figure 1-1 below).
4
See the next page to check the rating of the power-line fuse and replace with
the correct one if necessary.
5
Replace the cover and mark the power supply clearly with a tag or label
indicating the correct line voltage and fuse that is in use.
230V
100V
115V
Figure 1-1. Line Voltage Selector (set for 115Vac)
(TOP VIEW)
Summary of Contents for E364 A Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Page 28: ...2 General Information...
Page 40: ...3 Front Panel Operation and Features...
Page 68: ...4 Remote Interface Reference...
Page 114: ...5 Error Messages...
Page 124: ...6 Application Programs...
Page 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Page 136: ...7 Tutorial...
Page 147: ...Chapter 7 Tutorial Remote Programming 146...
Page 148: ...8 Specifications...
Page 156: ...Appendix Service Information...
Page 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Page 193: ...Appendix Service Information Replaceable Parts 192...
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......