128
Chapter 4
Troubleshooting Information
Checking PC Client Connections
E5810 appears
to be hung
There are several possible causes. For all of the following, use of client
timeouts and/or E5810 timeouts (the I/O timeout and LAN timeout
configuration values) can be used to ensure that the E5810 will not
“hang” indefinitely.
If you determine that the E5810 is truly hung, you may be able to
free it by terminating a particular client connection at the client,
thus freeing the resources to allow the E5810 to proceed.
Otherwise, you can press the
Clear ALL Pending Operations
button on the
Find and Control Instruments Connected to your
E5810
Web page or you can reboot the E5810.
The network connection to the E5810 has been broken.
The E5810 may be attempting a transaction to a non-existent
or powered-off device. Check the address used, as well as the
status of the device.
The E5810 is busy performing a very long transaction to a (slow)
device, or it is waiting for input from a device.
A client may be waiting for access to a device/interface locked
by another client. Check for possible deadlock situations and,
if possible, make sure the clients owning locks and the
connections to those clients are still up and functioning.
You can check the current session by clicking the
Session Status
icon to display the
E5810 LAN/GPIB Gateway Status
Web page.
Message
Description/Actions
Summary of Contents for Agilent E5810A
Page 1: ...Agilent E5810A LAN GPIB Gateway for Windows User s Guide ...
Page 2: ......
Page 6: ...Contents 6 ...
Page 15: ...1 E5810 Description ...
Page 33: ...2 Installing the E5810 ...
Page 75: ...3 Using E5810 Web Access ...
Page 85: ...Chapter 3 85 Using E5810 Web Access Viewing and Modifying Configuration ...
Page 108: ...108 Chapter 3 Using E5810 Web Access Other Web Access Functions ...
Page 109: ...4 Troubleshooting Information ...
Page 131: ...A E5810 Specifications ...
Page 136: ...136 AppendixA E5810 Specifications Supported Network Protocols Notes ...
Page 137: ...B Using the Agilent IO Libraries ...
Page 152: ...152 Appendix B Using the Agilent IO Libraries Installing Custom Agilent IO Libraries Notes ...
Page 153: ...C Using the Telnet Utility ...
Page 160: ...160 Appendix C Using the Telnet Utility Telnet Commands for the E5810 Notes ...
Page 161: ...161 Glossary ...
Page 172: ...172 Glossary Glossary Notes ...
Page 176: ...176 Index ...