538
Chapter 11
Synthesizer Section
Unlocked Reference PLL (100 MHz VCXO)
8. If the 100 MHz oscillator is working, the reason for the unlocked
condition is either a problem in the 10 MHz reference or a fault in
the signal path around the loop.
Check 10 MHz
reference to
phase/frequency
detector (steps
9-14)
9. On the spectrum analyzer, press
AUX CTRL
,
REAR PANEL
, and
10 MHz
INT
.
10.Check the 10 MHz reference frequency-accuracy by connecting a
frequency counter to A15J301 and verify that the reference
frequency is 10 MHz
±
40 Hz after a 5 minute warm-up period.
11.If a 10 MHz signal >1 V peak-to-peak is not present at A15J301,
refer to the
12.Measure the signal at TP301 with an oscilloscope. Refer to function
block M of A15 RF schematic (sheet 2 of 4).
13.Measure the signal at U502 pin 11 with an oscilloscope. Refer to
function block X of A15 RF schematic (sheet 2 of 4). This signal
should be TTL levels at 10 MHz with a 60 percent duty cycle.
14.If TTL-level signals (approximately 10 MHz) are not present, check
signals backwards through the loop to find a fault in the signal path.
15.Measure the signals at the following test points with an active
probe/spectrum analyzer combination:
16.If an approximately 10 MHz TTL signal is present at U502 pin 11
with 60 percent duty cycle, and the RF portion of the phase-lock loop
is functioning, the fault probably lies in the phase/frequency detector
or the 100 MHz lock loop integrator.
Check
phase/frequency
detector (steps
17-22)
17.Monitor U504 pin 5 and U503 pin 9 with an oscilloscope. These are
the two outputs of the phase/frequency detector. Refer to function
block O of A15 RF schematic (sheet 2 of 4).
18.A locked loop will exhibit stable, narrow (approximately 20 ns wide),
and positive-going TTL pulses occurring at a 10 MHz rate at U504
pin 5 and U503 pin 9.
Junction of C570 and C571
100 MHz, +2.5 dBm
±
2 dB
Junction of R715, R716,
100 MHz,
−
3 dBm
±
2 dB
R567, and R568
U700 pin 3
100 MHz, +16.5 dBm
±
2 dB
U700 pin 1
100 MHz, +8.5 dBm
±
2 dB
Summary of Contents for 8564EC
Page 17: ...25 1 General Information ...
Page 37: ...47 2 Adjustment Diagnostic Software ...
Page 77: ...89 3 Manual Adjustment Procedures ...
Page 129: ...161 3a Manual Adjustment Procedures 3335A Source not Available ...
Page 142: ...175 4 Assembly Replacement ...
Page 194: ...Chapter 4 257 Assembly Replacement Procedure 13 A21 OCXO Figure 4 34 A21 OCXO Mounting Screws ...
Page 196: ...259 5 Replaceable Parts ...
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ...303 6 Major Assembly and Cable Locations ...
Page 234: ...315 7 General Troubleshooting ...
Page 238: ...Chapter 7 319 General Troubleshooting Introduction Figure 7 2 Ribbon Cable Connections 1 of 3 ...
Page 239: ...320 Chapter7 General Troubleshooting Introduction Figure 7 3 Ribbon Cable Connections 2 of 3 ...
Page 242: ...Chapter 7 323 General Troubleshooting Introduction Figure 7 5 Service Cal Data Menu ...
Page 271: ...352 Chapter7 General Troubleshooting Block Diagram Description Figure 7 6 Functional Sections ...
Page 283: ...364 Chapter7 General Troubleshooting Block Diagram Description ...
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ...377 8 ADC Interface Section ...
Page 291: ...Chapter 8 381 ADC Interface Section Introduction Figure 8 2 A3 Test Connector Pin Locations ...
Page 321: ......
Page 322: ...427 9 IF Section ...
Page 356: ...Chapter 9 461 IF Section A5 IF Assembly Figure 9 13 Detailed IF Adjust Signature 5 ...
Page 376: ......
Page 377: ......
Page 378: ......
Page 379: ...487 10 Controller Section ...
Page 394: ...521 11 Synthesizer Section ...
Page 451: ......
Page 452: ......
Page 453: ......
Page 454: ...593 12 RF Section ...
Page 489: ...628 Chapter12 RF Section A15 RF Assembly Figure 12 10 10 MHz TTL Reference at U304 Pin 13 ...
Page 491: ......
Page 492: ...633 13 Display Power Supply Section ...
Page 504: ......
Page 505: ...671 14 Component Level Information Packets ...
Page 507: ......