Chapter 3
131
Manual Adjustment Procedures
5
. YTO Adjustment
Equipment
Microwave frequency counter ....................... 5343A Option 001
Adapters
Type N (m) to BNC (f) ................................................... 1250-1476
Type N (f) to 2.4 mm (f) ...................................................... 11903B
APC 3.5 (f) to APC 3.5 (f) .............................................. 5061-5311
Cables
BNC, 122 cm (48 in) ........................................................... 10503A
SMA, 61 cm (24 in) ........................................................ 8120-1578
Procedure
NOTE
This adjustment cannot be performed if preselected external mixer
mode is selected.
The
SAVELOCK ON OFF
function must be OFF.
YTO Main Coil Adjustments
1. Press
LINE
to turn the spectrum analyzer off. Remove the spectrum
analyzer cover and fold down the A15 RF and A14 frequency control
assemblies.
2. Disconnect the 50
Ω
termination from the 1ST LO OUTPUT. Connect
the equipment as shown in
LINE
to
turn the spectrum analyzer on.
3. Move the jumper on A14J23 from the NORM position (pins 1 and 2
jumpered) to the TEST position (pins 2 and 3 jumpered). See
3-14“YTO Adjustment Locations,” on page 132
for the location on the
A14 frequency control assemblies.
4. On the spectrum analyzer, press the following keys:
CONFIG
,
EXT MXR PRE UNPR
, (UNPR)
AUX CTRL
,
EXTERNAL MIXER
,
LOCK HARMONIC
, 6 Hz
SPAN
,
ZERO SPAN FREQUENCY
,
CENTER
FREQ
, 18.8893 GHz,
SGL SWP SAVE
,
SAVE STATE
,
STATE 0
FREQUENCY
, 35.7493 GHz
SAVE
,
SAVE STATE
,
STATE 1 RECALL
,
RECALL STATE
,
STATE 0
.
5. On the 5343A, press
SHIFT
7 and set the controls as follows:
Sample rate .......................................... Fully counterclockwise
10 Hz
−
500 MHz/500 MHz
−
26.5 GHz switch ......
500 MHz-26.5 GHz
Summary of Contents for 8564EC
Page 17: ...25 1 General Information ...
Page 37: ...47 2 Adjustment Diagnostic Software ...
Page 77: ...89 3 Manual Adjustment Procedures ...
Page 129: ...161 3a Manual Adjustment Procedures 3335A Source not Available ...
Page 142: ...175 4 Assembly Replacement ...
Page 194: ...Chapter 4 257 Assembly Replacement Procedure 13 A21 OCXO Figure 4 34 A21 OCXO Mounting Screws ...
Page 196: ...259 5 Replaceable Parts ...
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ...303 6 Major Assembly and Cable Locations ...
Page 234: ...315 7 General Troubleshooting ...
Page 238: ...Chapter 7 319 General Troubleshooting Introduction Figure 7 2 Ribbon Cable Connections 1 of 3 ...
Page 239: ...320 Chapter7 General Troubleshooting Introduction Figure 7 3 Ribbon Cable Connections 2 of 3 ...
Page 242: ...Chapter 7 323 General Troubleshooting Introduction Figure 7 5 Service Cal Data Menu ...
Page 271: ...352 Chapter7 General Troubleshooting Block Diagram Description Figure 7 6 Functional Sections ...
Page 283: ...364 Chapter7 General Troubleshooting Block Diagram Description ...
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ...377 8 ADC Interface Section ...
Page 291: ...Chapter 8 381 ADC Interface Section Introduction Figure 8 2 A3 Test Connector Pin Locations ...
Page 321: ......
Page 322: ...427 9 IF Section ...
Page 356: ...Chapter 9 461 IF Section A5 IF Assembly Figure 9 13 Detailed IF Adjust Signature 5 ...
Page 376: ......
Page 377: ......
Page 378: ......
Page 379: ...487 10 Controller Section ...
Page 394: ...521 11 Synthesizer Section ...
Page 451: ......
Page 452: ......
Page 453: ......
Page 454: ...593 12 RF Section ...
Page 489: ...628 Chapter12 RF Section A15 RF Assembly Figure 12 10 10 MHz TTL Reference at U304 Pin 13 ...
Page 491: ......
Page 492: ...633 13 Display Power Supply Section ...
Page 504: ......
Page 505: ...671 14 Component Level Information Packets ...
Page 507: ......