54
Chapter 2
Adjustment/Diagnostic Software
Getting Started
Getting Started
First, make sure you have a compatible controller (computer) and the
proper test equipment. The following paragraphs describe
requirements for controllers and test equipment. Once the proper
equipment is identified, proceed to
Spectrum Analyzers
The 8564E/8565E Adjustment/Diagnostic Software can be used to
adjust or troubleshoot ONLY an 8564EC or an 8565EC.
Errors will occur if this software is used to test spectrum analyzers
other than the 8564EC or 8565EC.
Controller (Computer)
The 8564E/8565E Adjustment/Diagnostic Software requires using one
of the following HP controllers and the HP BASIC operating system:
NOTE
When using an HP 9000 model 236 (HP 9836) controller, the graphics in
the IF diagnostics program will be very distorted and difficult to read.
The tabular data printed at the end of the IF diagnostics program
(which contains the same text information as the graphics) is clearly
readable.
Other HP 9000 Series 300 controllers are compatible with the
adjustment/diagnostic software; however, the graph printouts might
not be full width. This is especially true with medium- and
high-resolution displays.
The 8564E/8565E Adjustment/Diagnostic Software requires at least
4 Megabytes of free memory.
Controller
HP 9000 model 236 (HP 9836)
HP 9000 model 310
HP 9000 model 320
HP 9000 model 350
Operating System
HP BASIC 5.13 or later, and all BIN files
Summary of Contents for 8564EC
Page 17: ...25 1 General Information ...
Page 37: ...47 2 Adjustment Diagnostic Software ...
Page 77: ...89 3 Manual Adjustment Procedures ...
Page 129: ...161 3a Manual Adjustment Procedures 3335A Source not Available ...
Page 142: ...175 4 Assembly Replacement ...
Page 194: ...Chapter 4 257 Assembly Replacement Procedure 13 A21 OCXO Figure 4 34 A21 OCXO Mounting Screws ...
Page 196: ...259 5 Replaceable Parts ...
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ...303 6 Major Assembly and Cable Locations ...
Page 234: ...315 7 General Troubleshooting ...
Page 238: ...Chapter 7 319 General Troubleshooting Introduction Figure 7 2 Ribbon Cable Connections 1 of 3 ...
Page 239: ...320 Chapter7 General Troubleshooting Introduction Figure 7 3 Ribbon Cable Connections 2 of 3 ...
Page 242: ...Chapter 7 323 General Troubleshooting Introduction Figure 7 5 Service Cal Data Menu ...
Page 271: ...352 Chapter7 General Troubleshooting Block Diagram Description Figure 7 6 Functional Sections ...
Page 283: ...364 Chapter7 General Troubleshooting Block Diagram Description ...
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ...377 8 ADC Interface Section ...
Page 291: ...Chapter 8 381 ADC Interface Section Introduction Figure 8 2 A3 Test Connector Pin Locations ...
Page 321: ......
Page 322: ...427 9 IF Section ...
Page 356: ...Chapter 9 461 IF Section A5 IF Assembly Figure 9 13 Detailed IF Adjust Signature 5 ...
Page 376: ......
Page 377: ......
Page 378: ......
Page 379: ...487 10 Controller Section ...
Page 394: ...521 11 Synthesizer Section ...
Page 451: ......
Page 452: ......
Page 453: ......
Page 454: ...593 12 RF Section ...
Page 489: ...628 Chapter12 RF Section A15 RF Assembly Figure 12 10 10 MHz TTL Reference at U304 Pin 13 ...
Page 491: ......
Page 492: ...633 13 Display Power Supply Section ...
Page 504: ......
Page 505: ...671 14 Component Level Information Packets ...
Page 507: ......