124
Chapter 3
Manual Adjustment Procedures
3
. DC Log Amplifier Adjustments
4. Set up an 3335A as follows:
Frequency ..................................................................... 15 MHz
Amplitude ...................................................................
−
18 dBm
5. Press
CAL
,
ADJ CURR IF STATE
, wait for the analyzer to complete
adjustments then press
MKR
.
6. Adjust A4R445 for maximum on-screen amplitude. Refer to
A4 Linear Fidelity Adjustment
1. Press
LINE
to turn the spectrum analyzer off. Remove the spectrum
analyzer cover and place the spectrum analyzer in the service
position as illustrated in
. See
2. Connect the 3335A 50
Ω
output to the spectrum analyzer 50
Ω
input.
Press
LINE
to turn the spectrum analyzer on.
3. Press
PRESET AMPLITUDE
,
LINEAR
,
MORE 1 of 3
,
AMPTD UNITS
,
dBm
,
CAL
,
IF ADJ ON OFF
(OFF).
4. Press
PEAK SEARCH
,
MARKER DELTA
.
5. Reduce the 3335A input power to
−
58 dBm.
6. If the delta marker amplitude reads
−
40 dB
±
2 dB, no adjustment is
necessary.
7. If the signal is lower on the screen than expected ( delta marker
amplitude reads less than
−
42dB) then adjust A4R544 (see
) for an even lower level and press
CAL
,
ADJ CURR
IF STATE
. Allow sufficient time for the analyzer to complete the
adjustment.
8. If the signal is higher on the screen than expected ( delta marker
amplitude reads greater than
−
38 dB) then adjust A4R544 for an
even higher level signal and press
CAL
,
ADJ CURR IF STATE
. Allow
sufficient time for the analyzer to complete the adjustment.
9. Repeat steps 5 through 10.
A4 LOG Fidelity Adjustment
1. Press
LINE
to turn the spectrum analyzer off. Remove the spectrum
analyzer cover and place the spectrum analyzer in the service
position as illustrated in
. See
“DC Log Adjustment Locations,”
2. Connect the 3335 50
Ω
output to the spectrum analyzer 50
Ω
input.
Press
LINE
to turn the spectrum analyzer on.
Summary of Contents for 8564EC
Page 17: ...25 1 General Information ...
Page 37: ...47 2 Adjustment Diagnostic Software ...
Page 77: ...89 3 Manual Adjustment Procedures ...
Page 129: ...161 3a Manual Adjustment Procedures 3335A Source not Available ...
Page 142: ...175 4 Assembly Replacement ...
Page 194: ...Chapter 4 257 Assembly Replacement Procedure 13 A21 OCXO Figure 4 34 A21 OCXO Mounting Screws ...
Page 196: ...259 5 Replaceable Parts ...
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ...303 6 Major Assembly and Cable Locations ...
Page 234: ...315 7 General Troubleshooting ...
Page 238: ...Chapter 7 319 General Troubleshooting Introduction Figure 7 2 Ribbon Cable Connections 1 of 3 ...
Page 239: ...320 Chapter7 General Troubleshooting Introduction Figure 7 3 Ribbon Cable Connections 2 of 3 ...
Page 242: ...Chapter 7 323 General Troubleshooting Introduction Figure 7 5 Service Cal Data Menu ...
Page 271: ...352 Chapter7 General Troubleshooting Block Diagram Description Figure 7 6 Functional Sections ...
Page 283: ...364 Chapter7 General Troubleshooting Block Diagram Description ...
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ...377 8 ADC Interface Section ...
Page 291: ...Chapter 8 381 ADC Interface Section Introduction Figure 8 2 A3 Test Connector Pin Locations ...
Page 321: ......
Page 322: ...427 9 IF Section ...
Page 356: ...Chapter 9 461 IF Section A5 IF Assembly Figure 9 13 Detailed IF Adjust Signature 5 ...
Page 376: ......
Page 377: ......
Page 378: ......
Page 379: ...487 10 Controller Section ...
Page 394: ...521 11 Synthesizer Section ...
Page 451: ......
Page 452: ......
Page 453: ......
Page 454: ...593 12 RF Section ...
Page 489: ...628 Chapter12 RF Section A15 RF Assembly Figure 12 10 10 MHz TTL Reference at U304 Pin 13 ...
Page 491: ......
Page 492: ...633 13 Display Power Supply Section ...
Page 504: ......
Page 505: ...671 14 Component Level Information Packets ...
Page 507: ......