Chapter 11
537
Synthesizer Section
Unlocked Reference PLL (100 MHz VCXO)
Unlocked Reference PLL (100 MHz VCXO)
Operation (100 MHz VCXO)
The 600 MHz reference is generated by tripling, then doubling the
output of the 100 MHz phase-locked loop. If the 600 MHz reference is
off frequency, the 100 MHz phase-lock circuitry is probably at fault. If
there is no signal present at A15J701, or if the level is less than
−
3 dBm, the 100 MHz VCXO, the tripler, or the doubler circuitry has
probably failed. Refer to function blocks Q, R, and S of the A15 RF
schematic (sheet 2 of 4) in the 8560 E-Series Spectrum Analyzer
Component Level Information binder.
Troubleshooting (100 MHz VCXO)
Check 100 MHz
VCXO, tripler,
and doubler
(steps 1-7)
1. Using an active probe/spectrum analyzer combination, such as the
85024A/8566B, measure the tripler output at A15TP700. The tripler
output should be +3 dBm
±
2 dB.
2. If the tripler output is within tolerance, suspect the doubler
circuitry. Refer to function block S of the A15 RF schematic (sheet 2
of 4).
3. If the tripler output is too low, probe the output of A15U700 RF
amplifier. The level should be +16.5 dBm
±
2 dB. The level at the
input of A15U700 should be +8.5 dBm
±
2 dB.
4. If the level at the input of A15U700 is too low, suspect a faulty
100 MHz VCXO. Refer to function block Q of the A15 RF schematic
(sheet 2 of 4).
5. On the spectrum analyzer, press
AUX CTRL
,
REAR PANEL
, and
10 MHz INT
.
6. Measuring the tune voltage indicates if the 100 MHz PLL is locked.
Connect the ground lead the voltmeter to A15J1 pin 3 and measure
the voltage at A15J700 pin 3.
7. The tune voltage should be b1 and +24 Volts. If the tune
voltage is incorrect, place the P700 jumper (on A15J700) in the
TEST position (pin 1 to pin 2). This sets the tune voltage for varactor
A15CR700 to the n13 Volts, making it easier to troubleshoot
the 100 MHz VCXO, tripler, and doubler. Remember to return P700
jumper to the NORMAL position when you have finished
troubleshooting the oscillator circuitry.
Summary of Contents for 8564EC
Page 17: ...25 1 General Information ...
Page 37: ...47 2 Adjustment Diagnostic Software ...
Page 77: ...89 3 Manual Adjustment Procedures ...
Page 129: ...161 3a Manual Adjustment Procedures 3335A Source not Available ...
Page 142: ...175 4 Assembly Replacement ...
Page 194: ...Chapter 4 257 Assembly Replacement Procedure 13 A21 OCXO Figure 4 34 A21 OCXO Mounting Screws ...
Page 196: ...259 5 Replaceable Parts ...
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ...303 6 Major Assembly and Cable Locations ...
Page 234: ...315 7 General Troubleshooting ...
Page 238: ...Chapter 7 319 General Troubleshooting Introduction Figure 7 2 Ribbon Cable Connections 1 of 3 ...
Page 239: ...320 Chapter7 General Troubleshooting Introduction Figure 7 3 Ribbon Cable Connections 2 of 3 ...
Page 242: ...Chapter 7 323 General Troubleshooting Introduction Figure 7 5 Service Cal Data Menu ...
Page 271: ...352 Chapter7 General Troubleshooting Block Diagram Description Figure 7 6 Functional Sections ...
Page 283: ...364 Chapter7 General Troubleshooting Block Diagram Description ...
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ...377 8 ADC Interface Section ...
Page 291: ...Chapter 8 381 ADC Interface Section Introduction Figure 8 2 A3 Test Connector Pin Locations ...
Page 321: ......
Page 322: ...427 9 IF Section ...
Page 356: ...Chapter 9 461 IF Section A5 IF Assembly Figure 9 13 Detailed IF Adjust Signature 5 ...
Page 376: ......
Page 377: ......
Page 378: ......
Page 379: ...487 10 Controller Section ...
Page 394: ...521 11 Synthesizer Section ...
Page 451: ......
Page 452: ......
Page 453: ......
Page 454: ...593 12 RF Section ...
Page 489: ...628 Chapter12 RF Section A15 RF Assembly Figure 12 10 10 MHz TTL Reference at U304 Pin 13 ...
Page 491: ......
Page 492: ...633 13 Display Power Supply Section ...
Page 504: ......
Page 505: ...671 14 Component Level Information Packets ...
Page 507: ......