4 Connecting Mainframes to the Controller PC
Checking the Mainframe Clock Distribution
4-2
Agilent 81250 ParBERT Installation Guide, May 2004
Checking the Mainframe Clock
Distribution
After unpacking a mainframe, make sure that the clock distribution
cables have not loosened during transportation.
The
master clock distribution cables
connect the SYSTEM CLOCK
OUTPUTS of the clock module with the SYS CLK IN ports of the data
modules. They distribute the system master clock from the clock module
to the other modules of the clockgroup.
The figure below shows the cable installation of a system with an
E4805B/E4808A clock module. Black cables are used.
Figure 11 Arrangement of the Clock Distribution Cables (E4808A/E4805B Clock Modules)
The data modules for 7 and 13.5 Gbit/s require two clocks
―
the system
master clock and the
giga clock
. Both are provided by the E4809A clock
module.
E4805B/E4808A clock distribution
E4809A clock distribution