7 Special ParBERT Systems
Installing ParBERT 43/45G Systems
7-8
Agilent 81250 ParBERT Installation Guide, May 2004
To reconfigure the system to a ParBERT 43/45G:
1 Re-install the MUX or DEMUX module into the mainframe.
2 Connect the frontends to the MUX or DEMUX module.
The connection scheme is printed on the front panel of every
MUX/DEMUX module. The graphic shows eight consecutive data
modules with two frontends each.
0
2
4
6
8
10 12 14
1
3
5
7
9
11 13 15
Connection
Numbering
Figure 66 Numbering of Connections
As the ports of the MUX/DEMUX modules are numbered, this drawing
makes it very simple to establish the connections correctly:
–
Connect the two cables of port #0 of the MUX/DEMUX module to the
upper frontend of the leftmost data generator/analyzer module.
Connect “normal” to “normal” and “inverted” to “inverted”.
–
Connect port #1 of the MUX/DEMUX module to the lower frontend
of the leftmost data generator/analyzer module.
–
Connect port #2 of the MUX/DEMUX module to the upper frontend
of the next data generator/analyzer module.
–
Follow that scheme until all ports of the MUX/DEMUX module are
connected.
4 After power on, run the Agilent 81250 Configuration tool and activate
the “
Build Systems
” function.
Never attempt to operate an E4868B MUX or E4869B DEMUX module in
a general-purpose ParBERT system! These modules are highly susceptible
to overvoltage and electrostatic discharge.
ParBERT to ParBERT 45G conversion
CAUTION
Summary of Contents for 81250
Page 1: ...Agilent Parallel Bit Error Ratio Tester 81250 ParBERT Installation Guide Agilent Technologies ...
Page 12: ...1 About this Manual Document History 1 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 30: ...2 Introduction Unpacking ParBERT 2 18 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 139: ...Declaration of Conformity Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 3 ...
Page 141: ...Site Attenuation Requirements Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 5 ...
Page 142: ...Appendix Site Attenuation Requirements 10 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 145: ......