8 Troubleshooting the ParBERT Installation
Problems with Clockgroups
8-4
Agilent 81250 ParBERT Installation Guide, May 2004
Problem:
After adding modules or front-ends, the system can no longer run
synchronized tests.
Solution:
After changing any hardware in the system, it is absolutely necessary that
you deskew the system.
This consists of adjusting the instrument connectors and compensating
for internal and external delays. See
How to Adjust the Instrument
Connectors
and
How to Compensate for Internal and External Delays
in
the
Agilent 81250 ParBERT System User Guide
or in the online Help.
Problems with Clockgroups
Check the following if you have problems setting up clockgroups:
Problem:
For a multi-mainframe configuration, the software does not recognize the
master clockgroup.
Solution:
Have you rebuilt the systems?
When you change the configuration of the clock modules (for example,
when you set up one master clock module and one or two slave clock
modules), you have to reconfigure the systems in the software. See
to Configure I/O Interfaces and Systems” on page 5-9
for instructions.
Solution:
Are all the slave clock modules correctly connected to the master clock
module?
No synchronized tests
Master clockgroup not found
Summary of Contents for 81250
Page 1: ...Agilent Parallel Bit Error Ratio Tester 81250 ParBERT Installation Guide Agilent Technologies ...
Page 12: ...1 About this Manual Document History 1 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 30: ...2 Introduction Unpacking ParBERT 2 18 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 139: ...Declaration of Conformity Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 3 ...
Page 141: ...Site Attenuation Requirements Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 5 ...
Page 142: ...Appendix Site Attenuation Requirements 10 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 145: ......