Problems after Changing the Hardware
8 Troubleshooting the ParBERT Installation
Agilent 81250 ParBERT Installation Guide, May 2004
8-3
You may also check the ASCII file
dvtsys.txt
. This file informs you about
the systems that have been built and their module configuration.
Dvtsys.txt
is located in the subdirectory
\cfg
of the Agilent 81200
database.
In addition, you can run the VISA Resource Manager. Right-click the blue
IO
icon in the Windows task bar. Choose
Edit VXI Resource Manager
,
highlight the VXI interface, click
Run
, and finally click
RM Output
. The
VISA Resource Manager checks the mainframe configuration and reports
for each module:
•
Current VXI logical address (LADD)
•
Slot number
•
Manufacturer
•
Module identification
Most problems are caused by incorrect logical addresses. See
the Logical Address” on page 6-5
Problem:
ParBERT does not recognize newly inserted frontends.
Solution:
Have you rebuilt the systems?
Check the detected configuration with the Agilent 81250 Configuration
tool.
You may also check the file
dvtits.txt
in the database subdirectory
\cfg
.
It is recommended to open both files,
dvtsys.txt
and
dvtits.txt
, and to
inspect them in parallel.
Dvtsys.txt
defines the available systems,
dvtits.txt
provides the details.
NOTE
Unrecognized frontends
Summary of Contents for 81250
Page 1: ...Agilent Parallel Bit Error Ratio Tester 81250 ParBERT Installation Guide Agilent Technologies ...
Page 12: ...1 About this Manual Document History 1 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 30: ...2 Introduction Unpacking ParBERT 2 18 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 139: ...Declaration of Conformity Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 3 ...
Page 141: ...Site Attenuation Requirements Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 5 ...
Page 142: ...Appendix Site Attenuation Requirements 10 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 145: ......