Installing a Module
6 Changing the ParBERT Hardware
Agilent 81250 ParBERT Installation Guide, May 2004
6-5
How to Set the Logical Address
Each time you rearrange the modules of a mainframe, you also have to
make sure that they have the correct logical addresses.
Consult your own mainframe configuration table. Problems will arise, if a
new module is set to an existing or illegal address.
If you cannot find the mainframe configuration table, you can run the
VISA Resource Manager. This requires that the mainframe is connected
and switched on. Right-click the blue
IO
icon in the Windows task bar.
Choose
Edit VXI Resource Manager
, highlight the VXI interface, click
Run
, and finally click
RM Output
. The VISA Resource Manager checks the
mainframe configuration and reports for each module:
•
Current VXI logical address (LADD)
•
Slot number
•
Manufacturer
•
Module identification
When you are setting the logical address, ignore the numbers and
ON/OFF information on the DIP switch—refer to the information printed
on the module’s side cover.
7
LOGICAL ADDRESS
6
5
4
3
2
1
0
0
1
Figure 44 Module Address Switches
NOTE
TIP
Summary of Contents for 81250
Page 1: ...Agilent Parallel Bit Error Ratio Tester 81250 ParBERT Installation Guide Agilent Technologies ...
Page 12: ...1 About this Manual Document History 1 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 30: ...2 Introduction Unpacking ParBERT 2 18 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 139: ...Declaration of Conformity Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 3 ...
Page 141: ...Site Attenuation Requirements Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 5 ...
Page 142: ...Appendix Site Attenuation Requirements 10 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 145: ......