9 Controlling ParBERT Remotely
Controlling ParBERT via GPIB
9-2
Agilent 81250 ParBERT Installation Guide, May 2004
To operate ParBERT via GPIB:
1 Skip this step, if you have a preconfigured ParBERT controller PC.
Use the IO Config utility to configure the interface (see
Configure the IEEE 1394 and GPIB Interfaces” on page 5-2
Agilent IO Libraries Installation Guide
).
Make sure that you configure ParBERT as a listener. If necessary,
change the default GPIB settings.
2 Use the Agilent 81250 Configuration tool and set the startup mode of
ParBERT to “
controlled
“How to Specify Startup Settings” on
). The default GPIB address is 11.
3 Start the ParBERT user software.
In controlled mode, this starts the firmware server and the GPIB
gateway, not the graphical user interface.
If you wish to check or change GPIB parameters, please see the online
Help and refer to the
Agilent 81250 ParBERT System User Guide
.
For debugging purposes, it can be useful to enable the ParBERT user
interface even in controlled mode. This slows down the system’s
performance, but you can immediately check the reaction on your
commands.
TIP
Summary of Contents for 81250
Page 1: ...Agilent Parallel Bit Error Ratio Tester 81250 ParBERT Installation Guide Agilent Technologies ...
Page 12: ...1 About this Manual Document History 1 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 30: ...2 Introduction Unpacking ParBERT 2 18 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 139: ...Declaration of Conformity Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 3 ...
Page 141: ...Site Attenuation Requirements Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 5 ...
Page 142: ...Appendix Site Attenuation Requirements 10 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 145: ......