4.
Press
4
Meas
5 ,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
INST
TYPE
,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INST
TYPE:
VCO
TESTER
to
set
the
instrument
type
to
T
ester
Mode
.
5.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DC
POWER
CURRENT
to
set
the
measurement
item
to
DC
P
ower
Current.
6.
Press
4
DC
Pow
er
5 ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER
VOLTAGE
,
4
0
5 ,
4
x1
5
to
set
the
DC
POWER
voltage
to
0
V
.
7.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OUTPUT
on
OFF
to
output
the
DC
POWER
voltage
.
8.
Record
the
multimeter
reading
on
the
P
erformance
T
est
Record.
9.
Press
4
DC
Pow
er
5 ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER
VOLTAGE
,
and
the
entry
keys
to
set
the
DC
POWER
voltage
in
accordance
with
the
following
table
.
Subtract
each
DC
POWER
CURRENT
measurement
value
from
each
multimeter
reading,
and
record
the
calculated
value
on
the
P
erformance
T
est
Record.
DC
PO
WER
setting
voltage
DC
PO
WER
Current
through
300
R-Box
0.3
V
1
mA
3
V
10
mA
6
V
20
mA
9
V
30
mA
12
V
40
mA
15
V
50
mA
P
erformance
T
ests
2-37
Summary of Contents for 4352B
Page 11: ......
Page 31: ......
Page 73: ......
Page 80: ...Figure 3 4 List Box Menu Adjustments and Correction Constants 3 7 ...
Page 82: ...Figure 3 6 Reference Frequency Adjustment Location Adjustments and Correction Constants 3 9 ...
Page 85: ...Figure 3 9 Third IF AMP GAIN Adjustment Location 3 12 Adjustments and Correction Constants ...
Page 103: ...Figure 4 1 Troubleshooting Organization 4 2 Overall T roubleshooting ...
Page 113: ......
Page 130: ...Figure 5 11 Power Supply Block Diagram 1 Power Supply T roubleshooting 5 17 ...
Page 131: ...Figure 5 12 Power Supply Block Diagram 2 5 18 Power Supply T roubleshooting ...
Page 132: ...Figure 5 13 Power Supply Block Diagram 3 Power Supply T roubleshooting 5 19 ...
Page 133: ......
Page 135: ...Figure 6 1 Digital Control Group Simpli ed Block Diagram 6 2 Digital Control T roubleshooting ...
Page 146: ...Figure 7 2 Source Group Troubleshooting Flow Source Group T roubleshooting 7 3 ...
Page 154: ...Figure 8 2 Receiver Group Troubleshooting Flow Receiver Group T roubleshooting 8 3 ...
Page 198: ...Figure 10 1 Power Supply Functional Group Simpli ed Block Diagram Theory of Operation 10 3 ...
Page 203: ...Figure 10 3 Digital Control Group Simpli ed Block Diagram 10 8 Theory of Operation ...
Page 269: ......
Page 272: ...Figure B 1 Power Cable Supplied Power Requirement B 3 ...
Page 273: ......