A1
CPU
Replacement
When
you
replave
a
faulty
A1
CPU
with
a
new
one
,
remove
the
EEPROM
from
the
faulty
A1
and
mount
the
EEPROM
on
the
replacement
A1.
In
the
EEPROM,
the
correction
constants
data
is
stored
after
performing
the
A
djustment
and
Correction
Constants
procedures
described
in
the
chapter
2.
The
data
may
be
valid
for
the
new
A1
CPU
.
Figure
6-2.
A1
EEPROM
Location
Digital
Control
T
roubleshooting
6-3
Summary of Contents for 4352B
Page 11: ......
Page 31: ......
Page 73: ......
Page 80: ...Figure 3 4 List Box Menu Adjustments and Correction Constants 3 7 ...
Page 82: ...Figure 3 6 Reference Frequency Adjustment Location Adjustments and Correction Constants 3 9 ...
Page 85: ...Figure 3 9 Third IF AMP GAIN Adjustment Location 3 12 Adjustments and Correction Constants ...
Page 103: ...Figure 4 1 Troubleshooting Organization 4 2 Overall T roubleshooting ...
Page 113: ......
Page 130: ...Figure 5 11 Power Supply Block Diagram 1 Power Supply T roubleshooting 5 17 ...
Page 131: ...Figure 5 12 Power Supply Block Diagram 2 5 18 Power Supply T roubleshooting ...
Page 132: ...Figure 5 13 Power Supply Block Diagram 3 Power Supply T roubleshooting 5 19 ...
Page 133: ......
Page 135: ...Figure 6 1 Digital Control Group Simpli ed Block Diagram 6 2 Digital Control T roubleshooting ...
Page 146: ...Figure 7 2 Source Group Troubleshooting Flow Source Group T roubleshooting 7 3 ...
Page 154: ...Figure 8 2 Receiver Group Troubleshooting Flow Receiver Group T roubleshooting 8 3 ...
Page 198: ...Figure 10 1 Power Supply Functional Group Simpli ed Block Diagram Theory of Operation 10 3 ...
Page 203: ...Figure 10 3 Digital Control Group Simpli ed Block Diagram 10 8 Theory of Operation ...
Page 269: ......
Page 272: ...Figure B 1 Power Cable Supplied Power Requirement B 3 ...
Page 273: ......