Error
Messages
(Alphabetical
Order)
0350
Queue
overow
A
specic
code
entered
into
the
queue
instead
of
the
code
that
caused
the
error
.
This
code
indicates
that
there
is
no
room
in
the
queue
and
an
error
occurred
but
was
not
recorded.
0400
Query
error
This
is
the
generic
query
error
that
the
4352B
displays
when
it
cannot
detect
more
specic
errors
.
This
code
indicates
only
that
a
query
error
as
dened
in
IEEE
488.2,
11.5.1.1.7
and
6.3
has
occurred.
0410
Query
INTERRUPTED
A
condition
causing
an
interrupted
query
error
occurred
(see
IEEE
488.2,
6.3.2.3).
F
or
example
,
a
query
followed
by
DAB
or
GET
before
a
response
was
completely
sent.
0420
Query
UNTERMINA
TED
A
condition
causing
an
unterminated
query
error
occurred
(see
IEEE
488.2,
6.3.2.2).
F
or
example
,
the
4352B
was
addressed
to
talk
and
an
incomplete
program
message
was
receved
by
the
controller
.
129
RECALL
ERROR:
INSTR
ST
A
TE
PRESET
A
serious
error
,
for
example
corrupted
data,
is
detected
on
recalling
a
le
,
and
this
forced
the
4352B
to
be
PRESET
.
49
RF
Freq
Out
Of
Range
The
DUT's
output
frequency
is
not
within
the
measurement
range
of
the
4352B
(10
MHz
to
3
GHz).
V
erify
the
DUT's
output
frequency
.
58
RF
Input
Over
Trans
Max
Freq
Indicates
that
the
frequency
transient
measurement
trace
was
probably
inaccurate
because
the
input
frequency
to
the
4352B
was
above
the
upper
limit
of
the
detection
band.
Note
,
however
,
that
\RF
Input
Under
Trans
Min
Freq "
(message
No
.:
57)
may
erroneously
appear
even
when
the
above
problem
is
present,
if
the
4352B
is
operating
in
the
heterodyne
mode
and
if
the
dierence
between
input
signals
from
the
device
and
the
external
signal
generator
is
larger
than
40
MHz.
V
erify
that
the
DUT
generates
the
proper
signal,
and
sends
it
to
the
4352B
RF
IN
connector
.
If
the
signal
is
generated,
check
its
frequency
.
Press
4
Sense
Range
5 ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MAX
TRANS
FREQ
or
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MIN
TRANS
FREQ
,
and
the
entry
keys
to
change
the
detection
bandwidth,
if
required.
57
RF
Input
Under
Trans
Min
Freq
Indicates
that
the
frequency
transient
measurement
trace
was
probably
inaccurate
because
the
input
frequency
to
the
4352B
was
below
the
lower
limit
of
the
detection
band.
Note
,
however
,
that
\RF
Input
Under
Trans
Max
Freq "
(message
No
.:
58)
may
erroneously
appear
even
when
the
above
problem
is
present,
if
the
4352B
is
operating
in
the
heterodyne
mode
and
if
the
dierence
between
input
signals
from
the
device
and
the
external
signal
generator
is
larger
than
40
MHz.
Messages-11
Summary of Contents for 4352B
Page 11: ......
Page 31: ......
Page 73: ......
Page 80: ...Figure 3 4 List Box Menu Adjustments and Correction Constants 3 7 ...
Page 82: ...Figure 3 6 Reference Frequency Adjustment Location Adjustments and Correction Constants 3 9 ...
Page 85: ...Figure 3 9 Third IF AMP GAIN Adjustment Location 3 12 Adjustments and Correction Constants ...
Page 103: ...Figure 4 1 Troubleshooting Organization 4 2 Overall T roubleshooting ...
Page 113: ......
Page 130: ...Figure 5 11 Power Supply Block Diagram 1 Power Supply T roubleshooting 5 17 ...
Page 131: ...Figure 5 12 Power Supply Block Diagram 2 5 18 Power Supply T roubleshooting ...
Page 132: ...Figure 5 13 Power Supply Block Diagram 3 Power Supply T roubleshooting 5 19 ...
Page 133: ......
Page 135: ...Figure 6 1 Digital Control Group Simpli ed Block Diagram 6 2 Digital Control T roubleshooting ...
Page 146: ...Figure 7 2 Source Group Troubleshooting Flow Source Group T roubleshooting 7 3 ...
Page 154: ...Figure 8 2 Receiver Group Troubleshooting Flow Receiver Group T roubleshooting 8 3 ...
Page 198: ...Figure 10 1 Power Supply Functional Group Simpli ed Block Diagram Theory of Operation 10 3 ...
Page 203: ...Figure 10 3 Digital Control Group Simpli ed Block Diagram 10 8 Theory of Operation ...
Page 269: ......
Page 272: ...Figure B 1 Power Cable Supplied Power Requirement B 3 ...
Page 273: ......