![Agilent Technologies 4352B Service Manual Download Page 139](http://html.mh-extra.com/html/agilent-technologies/4352b/4352b_service-manual_2867893139.webp)
Check
Error
Messages
Turn
the
analyzer
power
on.
Check
that
no
error
message
appears
on
the
LCD
.
If
no
error
message
is
displayed,
continue
with
the
Check
A1
DRAM
and
Flash
Memory
in
this
ST
ART
HERE
section.
If
one
of
error
messages
listed
below
is
displayed,
follow
the
instruction
described
below
.
F
or
any
other
message
,
see
the
Error
Messages
in
the
Messages
Chapter
.
Error
Messages
Instruction
EEPROM
CHECK
SUM
ERROR
This
indicates
that
the
correction
constants
stored
in
the
EEPROM
on
the
A1
CPU
are
invalid
or
the
EEPROM
is
faulty
.
Rewrite
all
correction
constants
into
the
EEPROM
using
the
adjustment
program.
F
or
the
detailed
procedure
,
see
the
A
djustments
and
Correction
Constants
chapter
in
this
manual.
If
the
rewriting
is
not
successfully
performed,
replace
the
EEPROM
and
then
rewrite
the
correction
constants
into
the
new
EEPROM.
Svc
(Status
Annotation)
This
indicates
that
the
correction
constants
stored
in
the
EEPROM
on
the
A1
CPU
are
invalid
or
the
EEPROM
is
faulty
.
See
the
instructions
in
the
EEPROM
CHECK
SUM
ERROR
message
,
above
.
CPU
DRAM
R/W
ERROR
This
indicates
that
the
A1
DRAM
Read/Write
test
failed
on
the
power-on
self
test
sequence
.
Replace
the
A1
CPU
.
DUAL
PORT
SRAM
R/W
ERROR
This
indicates
that
the
power-on
self
test
for
DSP's
dual
port
SRAM
on
the
A1
CPU
failed.
Replace
the
A1
CPU
.
Note
CAUTION:
Insufficient
RF
Level
is
displayed
when
turning
with
the
analyzer
power
on.
This
is
caused
by
no
RF
signal
input
on
RF
IN
connector
of
the
4352B.
After
the
signal
is
input
to
the
RF
IN
connector
,
this
message
will
disappear
.
2.
Check
the
A1
CPU
a.
Turn
the
analyzer
power
on.
b.
Press
4
System
5 ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
MENU
,
NNNNNNNNNNNNNNNNN
TESTS
,
4
1
5 ,
4
x1
5 ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXECUTE
TEST
to
run
the
internal
test
1:
A1
CPU
.
If
no
error
message
is
displayed,
the
A1
CPU
is
probably
good.
Continue
with
the
next
step
,
Check
the
A1
DRAM
and
Flask
Memory
.
If
one
of
the
error
messages
listed
below
is
displayed,
the
A1
CPU
is
faulty
.
Replace
the
A1
CPU
.
FLASH
MEMORY
CHECK
SUM
ERROR
EEPROM
CHECK
SUM
ERROR
F-BUS
TIMER
CHIP
TEST
FAILED
KEY
CHIP
TEST
FAILED
GPIB
CHIP
TEST
FAILED
RTC
CHIP
TEST
FAILED
mini-DIN
CHIP
TEST
FAILED
DSP
CHIP
TEST
FAILED
BACKUP
SRAM
CHECK
SUM
ERROR
FDC
CHIP
TEST
FAILED
6-6
Digital
Control
T
roubleshooting
Summary of Contents for 4352B
Page 11: ......
Page 31: ......
Page 73: ......
Page 80: ...Figure 3 4 List Box Menu Adjustments and Correction Constants 3 7 ...
Page 82: ...Figure 3 6 Reference Frequency Adjustment Location Adjustments and Correction Constants 3 9 ...
Page 85: ...Figure 3 9 Third IF AMP GAIN Adjustment Location 3 12 Adjustments and Correction Constants ...
Page 103: ...Figure 4 1 Troubleshooting Organization 4 2 Overall T roubleshooting ...
Page 113: ......
Page 130: ...Figure 5 11 Power Supply Block Diagram 1 Power Supply T roubleshooting 5 17 ...
Page 131: ...Figure 5 12 Power Supply Block Diagram 2 5 18 Power Supply T roubleshooting ...
Page 132: ...Figure 5 13 Power Supply Block Diagram 3 Power Supply T roubleshooting 5 19 ...
Page 133: ......
Page 135: ...Figure 6 1 Digital Control Group Simpli ed Block Diagram 6 2 Digital Control T roubleshooting ...
Page 146: ...Figure 7 2 Source Group Troubleshooting Flow Source Group T roubleshooting 7 3 ...
Page 154: ...Figure 8 2 Receiver Group Troubleshooting Flow Receiver Group T roubleshooting 8 3 ...
Page 198: ...Figure 10 1 Power Supply Functional Group Simpli ed Block Diagram Theory of Operation 10 3 ...
Page 203: ...Figure 10 3 Digital Control Group Simpli ed Block Diagram 10 8 Theory of Operation ...
Page 269: ......
Page 272: ...Figure B 1 Power Cable Supplied Power Requirement B 3 ...
Page 273: ......