T
est
Result
(dB)
=
4352B
reading
-
(P
ower
Meter
reading
-
Track(50M))
Record
the
calculated
test
result
on
the
P
erformance
T
est
Record.
16.
Turn
the
signal
generator
output
OFF
,
and
set
the
signal
generator
output
level
to
-12
dBm,
50
MHz.
Then
turn
the
signal
generator
output
ON.
17.
A
djust
the
signal
generator
output
level
until
the
power
meter
reading
is
10dBm
60.1
dB
.
18.
Read
the
4352B
measurement
result,
and
calculate
the
following
formula:
T
est
Result
(dB)
=
4352B
reading
-
(P
ower
Meter
reading
-
Track(50M))
Record
the
calculated
test
result
on
the
P
erformance
T
est
Record.
Caution
T
o
prevent
damage
of
the
4352B
from
an
accidentially
large
signal
level
input,
turn
the
power
amplier
o
immediately
after
this
test
is
completed.
2-10
P
erformance
T
ests
Summary of Contents for 4352B
Page 11: ......
Page 31: ......
Page 73: ......
Page 80: ...Figure 3 4 List Box Menu Adjustments and Correction Constants 3 7 ...
Page 82: ...Figure 3 6 Reference Frequency Adjustment Location Adjustments and Correction Constants 3 9 ...
Page 85: ...Figure 3 9 Third IF AMP GAIN Adjustment Location 3 12 Adjustments and Correction Constants ...
Page 103: ...Figure 4 1 Troubleshooting Organization 4 2 Overall T roubleshooting ...
Page 113: ......
Page 130: ...Figure 5 11 Power Supply Block Diagram 1 Power Supply T roubleshooting 5 17 ...
Page 131: ...Figure 5 12 Power Supply Block Diagram 2 5 18 Power Supply T roubleshooting ...
Page 132: ...Figure 5 13 Power Supply Block Diagram 3 Power Supply T roubleshooting 5 19 ...
Page 133: ......
Page 135: ...Figure 6 1 Digital Control Group Simpli ed Block Diagram 6 2 Digital Control T roubleshooting ...
Page 146: ...Figure 7 2 Source Group Troubleshooting Flow Source Group T roubleshooting 7 3 ...
Page 154: ...Figure 8 2 Receiver Group Troubleshooting Flow Receiver Group T roubleshooting 8 3 ...
Page 198: ...Figure 10 1 Power Supply Functional Group Simpli ed Block Diagram Theory of Operation 10 3 ...
Page 203: ...Figure 10 3 Digital Control Group Simpli ed Block Diagram 10 8 Theory of Operation ...
Page 269: ......
Page 272: ...Figure B 1 Power Cable Supplied Power Requirement B 3 ...
Page 273: ......