Agilent 1260 Infinity II Preparative Autosampler User Manual
5
Preparing the Module
Swage Fittings
5
Once you believe you have the fitting complete, loosen
the nut, and inspect the ferrule for the correct position
on the tubing.
Example of a perfect fitting
Examples of incorrect fittings
Figure 16
Examples of incorrect fittings
0L[LQJFKDPEH
OHDNVZLOORFFXU
)HUUXOHFDQQRWVHDWSURSHUO\
,IGLPHQVLRQ;LVWRRORQJ
,IGLPHQVLRQ;LVWRRVK
DGHDGYROXPHRU
PL[LQJFKDPEHUZLOORF
;
;
N O T E
The first time that the swagelock fitting is used on a column or an injection valve, the
position of the ferrule is permanently set. If changing from a column or an injection valve
to another, the fitting may leak or decrease the quality of the separation by contributing to
band broadening.