Agilent 1260 Infinity II Preparative Autosampler User Manual
9
Maintenance
Exchange the Gripper Arm
5
Rotate the gripper arm slowly from left to right and apply
a gentle pressure to the paper clip.
The clip will engage on an internal catch and the rotation
of the arm will be blocked.
6
Hold the paper clip in place, press the gripper release
button and rotate the gripper arm to the right.
7
The gripper arm will come off.
8
Replace the gripper arm by holding the paper clip in
place, pushing the gripper arm into the holder and
rotating the gripper arm to the left.
9
In the Local Controller close
Change gripper
.
OR
In the Agilent Lab Advisor software select
Change
gripper > Next
, click
Next
and wait until the gripper is in
the default position.
10
Select
Back
to finish the procedure.
11
Close the doors.