Agilent 1260 Infinity II Preparative Autosampler User Manual
8
Error Information
Sampler Error Messages
Initialization Failed
Error ID: 34392, 34391
The sampler failed to complete initialization correctly.
The sampler initialization procedure moves the needle arm and transport
assembly to their home positions in a predefined sequence. During
initialization, the processor monitors the position sensors and motor encoders
to check for correct movement. If one or more of the movements is not
successful, or is not detected, the error message is generated.
Probable cause
Suggested actions
1
Mechanical obstruction.
Ensure unobstructed movement of the
transport assembly.
2
Defective needle station board.
Exchange and reset the needle station board.
3
Defective transport assembly flex board.
Exchange the transport assembly.
4
Defective hydraulic unit motor.
Exchange the hydraulic unit motor.
5
Defective main board.
Exchange the main board.