![Agilent Technologies 1260 Infinity II User Manual Download Page 134](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii/1260-infinity-ii_user-manual_2867537134.webp)
134
Agilent 1260 Infinity II Preparative Autosampler User Manual
8
Error Information
General Error Messages
Leak Sensor Short
Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit,
the error message is generated.
Compensation Sensor Open
Error ID: 0081
The ambient-compensation sensor (NTC) on the power switch board in the
module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
power switch board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient temperature
changes. If the resistance across the sensor increases above the upper limit,
the error message is generated.
Probable cause
Suggested actions
1
Defective leak sensor.
Exchange the leak sensor.
2
Leak sensor incorrectly routed, being
pinched by a metal component.
•
Make sure the leak sensor is installed
correctly.
•
Correct the routing of the cable.
•
If cable defective, exchange the leak
sensor.
3
Power switch assembly defective
Exchange the power switch assembly.
Probable cause
Suggested actions
1
Loose connection between the power
switch board and the main board
Check the ribbon cable between the power
switch board and the main board.
2
Defective power switch assembly
Exchange the power switch assembly.