Agilent 1260 Infinity II Preparative Autosampler User Manual
9
Maintenance
Exchange the Metering Seal and Piston
3
Disconnect the two capillaries from the metering device.
4
Remove the two fixing bolts.
5
Pull the metering head assembly away from the sampler.
6
Remove the two fixing bolts from the base of the
metering head assembly.
7
Disassemble the metering head assembly.
8
Use a seal insert tool to carefully remove the seal.