Agilent 1260 Infinity II Preparative Autosampler User Manual
31
Site Requirements and Specifications
2
Performance Specifications
Instrument Control
LC & CE Drivers A.02.17 or above
Instrument Control Framework (ICF) A.02.04 or
above
Instant Pilot (G4208A) with firmware B.02.22 or
above
Lab Advisor B.02.10 or above
For details about supported software versions
refer to the compatibility matrix of your version
of the LC and CE Drivers
Communications
Controller-area network (CAN), Local Area
Network (LAN)
ERI: ready, start, stop and shut-down signals
Safety and
maintenance
Extensive diagnostics, error detection and
display with Agilent Lab Advisor software
Leak detection, safe leak handling, leak output
signal for shutdown of pumping system, and low
voltages in major maintenance areas
GLP features
Early maintenance feedback (EMF) for
continuous tracking of instrument usage with
user-settable limits and feedback messages.
Electronic records of maintenance and errors.
Housing
All materials recyclable.
Table 2
Performance Specifications 1260 Infinity II Preparative Autosampler (G7157A)
Type
Specification
Comment