164
Agilent 1260 Infinity II Preparative Autosampler User Manual
9
Maintenance
Exchange the Sample Loop Assembly
Exchange the Sample Loop Assembly
When
When sample loop is visibly damaged, blocked or leaks.
Tools required
p/n
Description
8710-1924
Wrench open 14 mm
Parts required
p/n
Description
G7157-60515
Loop for Prep Autosampler
Preparations
Finish any pending acquisition job and in order to avoid leaks, stop the pump running and remove
the tubings from the solvent bottles. If available close the shutoff valves.
WA R N I N G
Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
➔
Do not open the safety cover of the needle station during normal operation.
➔
Wear safety goggles and safety gloves when removing the needle assembly.
1
In the Instant Pilot start the maintenance mode and
select
Change loop
function.
OR
In the Agilent Lab Advisor software select
Service &
Diagnostics
in the system screen
Maintenance
Positions > Change loop
, click
Start
and wait until the
needle assembly is in maintenance position.
2
Open the front door.