Agilent 1260 Infinity II Preparative Autosampler User Manual
33
Site Requirements and Specifications
2
Specifications of the Sample Cooler
C A U T I O N
General hazards and improper disposal
Improper disposal of the media and components used pollutes the environment.
➔
The breakdown of the sample cooler unit must be carried out by specialist
refrigeration company.
➔
All media must be disposed of in accordance with national and local regulations.
➔
Please contact your local Agilent Service Center in regard to safe environmental
disposal of the appliance or check
www.agilent.com
for more info.
Table 4
Performance Specifications Agilent 1290 Sample Cooler
Type
Specifications
Operating principle
High performance, low-energy consumption
micro-compressor based cooler with
ozone-friendly HFC-134a coolant (42 g),
user-upgradable.
Temperature range
from 4 °C to 5 °C below ambient
Temperature settable
from 4 – 40 °C in 1 ° increments
Temperature accuracy (<25 °C, <50 % r.H.)
2 °C to 6 °C at a setpoint of 4 °C