Agilent 1260 Infinity II Preparative Autosampler User Manual
7
Troubleshooting and Diagnostics
Maintenance Functions
Change Gripper
The change gripper function moves the gripper to the front of the autosampler
enabling easy access to the gripper release mechanism.
User Interface
The commands for the control software are:
Start
Moves the transport assembly and gripper arm to the position required to
change the gripper arm.
Next
Repositions the transport assembly and gripper arm to the home position.
Back
Completes the procedure.
Using the
Change Gripper
Function
1
Select
Start
to move the gripper arm to the maintenance position.
2
3
Select
Next
to move the gripper arm to the home position.
4
Select
Back
to finish the gripper procedure.
N O T E
Read carefully the description in the LabAdvisor user interface.