Agilent InfinityLab LC Series RID User Manual
163
12
Hardware Information
Firmware Description
Firmware Updates
Firmware updates can be done with the Agilent Lab Advisor software with files
on the hard disk (latest version should be used).
Required tools, firmware and documentation are available from the Agilent web:
http://www.agilent.com/en-us/firmwareDownload?whid=69761
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
•
PPPP is the product number, for example, 1315B for the G1315B DAD,
•
R the firmware revision, for example, A for G1315B or B for the G1315C DAD,
•
VVV is the revision number, for example 650 is revision 6.50,
•
XXX is the build number of the firmware.
For instructions on firmware updates refer to section
Replacing Firmware
in
chapter
"Maintenance"
or use the documentation provided with the
Firmware
Update Tools
.
Figure 27
Firmware Update Mechanism
NOTE
Update of main system can be done in the resident system only. Update of the
resident system can be done in the main system only.
Main and resident firmware must be from the same set.
Resident System
Main FW update
Main System
Resident FW Update