Lost CAN Partner
Error ID: 0071
During an analysis, the internal synchronization or communication between one
or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or
more of the modules is no longer recognized as being connected to the system,
the error message is generated.
Leak Sensor Open
Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the error
message is generated.
Probable cause
Suggested actions
1
CAN cable disconnected.
•
Ensure all the CAN cables are connected cor-
rectly.
•
Ensure all CAN cables are installed correctly.
2
Defective CAN cable.
Exchange the CAN cable.
3
Defective mainboard in another module.
Switch off the system. Restart the system, and
determine which module or modules are not recog-
nized by the system.
Probable cause
Suggested actions
1
Leak sensor not connected to the Power
Switch board.
Please contact your Agilent service representative.
2
Defective leak sensor.
Please contact your Agilent service representative.
3
Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service representative.
4
Power switch assembly defective
Please contact your Agilent service representative.