Agilent InfinityLab LC Series RID User Manual
140
9
Maintenance and Repair
Replace the Module Firmware
Replace the Module Firmware
To upgrade/downgrade the module’s firmware carry out the following steps:
1
Download the required module firmware, the latest FW Update Tool and the
documentation from the Agilent web.
http://www.agilent.com/en-us/firmwareDownload?whid=69761
2
For loading the firmware into the module follow the instructions in the
documentation.
Module Specific Information
When
The installation of newer firmware might be necessary
•
if a newer version solves problems of older versions or
•
to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
•
to keep all systems on the same (validated) revision or
•
if a new module with newer firmware is added to a system or
•
if third party control software requires a special version.
Tools required
Description
Agilent Lab Advisor software
Parts required
#
Description
1
Firmware, tools and documentation from Agilent web site
Preparations
Read update documentation provided with the Firmware Update Tool.
G7162A/B
Initial firmware
D.06.76
Compatibility with 1100 / 1200 series modules
When using the G7162A/B in a system, all other
modules must have firmware from set 6.50 or
above (main and resident). Otherwise the com-
munication will not work.
Conversion to / emulation
Not available