Agilent InfinityLab LC Series RID User Manual
170
12
Hardware Information
Interfaces
Overview Interfaces
CAN
The CAN is inter-module communication interface. It is a 2-wire serial bus
system supporting high speed data communication and real-time requirement.
LAN
The modules have either an interface slot for a LAN card (e.g. Agilent G1369B/C
LAN Interface) or they have an on-board LAN interface (e.g. detectors G1315C/D
DAD and G1365C/D MWD). This interface allows the control of the
module/system via a PC with the appropriate control software. Some modules
have neither on-board LAN nor an interface slot for a LAN card (e.g. G1170A
Valve Drive or G4227A Flexible Cube). These are hosted modules and require a
Host module with firmware B.06.40 or later or with additional G1369C LAN Card.
USB
The USB interface replaces the RS-232 Serial interface in new FUSION generation
modules. For details on USB refer to
Analog Signal Output
The analog signal output can be distributed to a recording device. For details
refer to the description of the module’s mainboard.
NOTE
If an Agilent detector (DAD/MWD/FLD/VWD/RID) is in the system, the LAN
should be connected to the DAD/MWD/FLD/VWD/RID (due to higher data load).
If no Agilent detector is part of the system, the LAN interface should be installed
in the pump or autosampler.