Agilent InfinityLab LC Series RID User Manual
88
7
Error Information
General Error Messages
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 4 of the remote connector generates the error
message.
Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one minute
of starting the analysis. If a not-ready condition is still present on the remote line
after one minute the error message is generated.
Probable cause
Suggested actions
1
Leak detected in an external instrument with
a remote connection to the system.
Fix the leak in the external instrument before
restarting the module.
2
Shut-down in an external instrument with a
remote connection to the system.
Check external instruments for a shut-down con-
dition.
3
The degasser failed to generate sufficient
vacuum for solvent degassing.
Check the vacuum degasser for an error condi-
tion. Refer to the
Service Manual
for the degasser
or the pump that has the degasser built-in.
Probable cause
Suggested actions
1
Not-ready condition in one of the instru-
ments connected to the remote line.
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up cor-
rectly for analysis.
2
Defective remote cable.
Exchange the remote cable.
3
Defective components in the instrument
showing the not-ready condition.
Check the instrument for defects (refer to the
instrument’s documentation).