Agilent InfinityLab LC Series RID User Manual
135
9
Maintenance and Repair
Correct Leaks
Correct Leaks
When
If a leakage has occurred in the valve area or at the capillary connections
Tools required
Description
Tissue
Wrench, 1/4 inch
for capillary connections
1
Open the doors.
2
Observe the leak sensor area for leaks and correct, if
required.
3
Observe the interface ports for leaks and correct, if
required.
4
Unscrew the screws of the Service Door.