Agilent InfinityLab LC Series RID User Manual
113
8
Test Functions
Refractive Index Calibration
4
If the detector response differs from the theoretical response of
512000 nRIU
±
5000 nRIU enter the theoretical value (512000) in the dialog
box. If the detector response is within the theoretical response click
OK
.
5
The calibration was successful.
NOTE
Rinse the sample cell with pure water at a minimum of 1.5 mL/min to flush the
sucrose from the cell and the capillaries. When organic solvent is sequentially
applied (without rinsing), a blockage of capillaries may occur.