Agilent InfinityLab LC Series RID User Manual
40
2
Site Requirements and Specifications
Performance Specifications
Performance Specifications
Table 3
Agilent 1290 Infinity II Refractive Index Detector (G7162B) Performance Specifications
Type
Specification
Detection type
Refractive Index
Refractive index range
1.00 – 1.75 RIU, calibrated
Measurement range
±
600·10
-6
RIU
Optical zeroing
Digitally via software, manually via set screw
Optics temperature control
5 °C above ambient to 55 °C
Sample cell
Volume: 2.5 µL
Maximum pressure: 5 bar (0.5 MPa)
Maximum flow rate: 1 mL/min (100 % water)
Valves
Automatic purge and automatic solvent recycle
Volumes
Inlet port to sample cell 2.5 µL,
inlet port to outlet port 265 µL
Liquid contact materials
316 stainless steel, PTFE
and quartz glass
pH range
2.3 – 9.5
Performance specifications
Short term noise:
<±
1.75·10
-9
RIU
Drift:
<
200·10
-9
RIU/hr
Time programmable param-
eters
Polarity, peak width
Maximum data rate
148 Hz
Detector zero
Automatic zero before analysis
Instrument Control
Lab Advisor B.02.07 or above
LC and CE Drivers A.02.12 or above
Local control
Agilent Instant Pilot (G4208A)
Analog outputs
Recorder/integrator: 100 mV or 1 V, output range selectable, one out-
put
Communication
LAN, Controller Area Network (CAN), Extended Remote Interface
(ERI): ready, start, stop and shut-down signals
Safety and maintenance
Extensive diagnostics, error detection and display, leak detection,
safe leak handling, leak output signal for shutdown of pumping sys-
tem. Low voltages in major maintenance areas.