DX-D 600
User Manual
0242F EN 20181218
141
SECTION 8
TROUBLESHOOTING GUIDE
A guide for a quick solution of main typical problems in the use of this equipment
follows. It is recommended to keep this troubleshooting guide with you when
operating with the equipment.
8.1
RAD WALL STAND DR
PROBLEM
CHECK IF
ACTION
Emergency Stop Switch is activated.
Deactivate Emergency Stop Switch.
RAD WALL STAND CAN NOT BE
SWITCHED ON
There is not enough power.
Check that the Line Power is provided to the RAD
Wall Stand from the RAD Table and the Room
Electrical Cabinet. If it is correct and it can not be
turned ON, contact Service.
There is any obstacle on the vertical
travel.
Remove anything that may obstruct the vertical
movement of the column stand.
DR DETECTOR VERTICAL
MOVEMENT IS NOT POSSIBLE
Press and hold the Vertical Lock Handle.
MOVEMENT IS NOT POSSIBLE
It is locked.
In case that Vertical lock is broken, contact
Service.
THE DR DETECTOR DOES NOT TILT
Power supply is OFF.
Switch on the System.
AUTO-TRACKING FUNCTIONALITY
DOES NOT WORK
Check the Ceiling Suspension Control
Console Error Message.
Complete the recommended Action by the Ceiling
Suspension Control Console.
8.2
RAD TABLE
PROBLEM
CHECK IF
ACTION
Emergency Stop Switch is activated.
Deactivate Emergency Stop Switch.
RAD TABLE CAN NOT BE
SWITCHED ON
There is not enough power.
Check that the Line Power is provided to the RAD
Table from the Room Electrical Cabinet. If it is
correct and it can not be turned ON, contact
Service.
There is not enough power.
Check conditions for ”RAD Table can not be
switched ON” above.
VERTICAL MOVEMENTS ARE
BLOCKED
Control Pedals are not working properly.
Verify that the Control Pedals are not blocked with
any obstacle, when the Pedals are pressed.
If it is correct and movement is not possible,
contact Service.
Anti-collision Switches do not work.
Contact Service.
Summary of Contents for DX-D 600
Page 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Page 2: ......
Page 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Page 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Page 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Page 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Page 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Page 189: ......