Table 8-3
Wall Stand Motor Controller Boards
CODE
USER ERROR MESSAGE
ERROR DESCRIPTION
ACTION
11201
Control Board Error:
Wall Stand-Vertical
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Replace the Control Driver Board
11202
Control Board Error:
Wall Stand-Tilting
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Replace the Control Driver Board
11204
Invalid calibration data:
Wall Stand-Vertical
Check Vertical Axis calibration
11205
Invalid calibration data:
Wall Stand-Tilting
Check Calibration
11210
Position out of range:
Wall Stand-Vertical
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check Vertical Axis calibration
11211
Control Board Error:
Wall Stand-Vertical
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Replace the Control Driver Board
Summary of Contents for DX-D 600
Page 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Page 2: ......
Page 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Page 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Page 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Page 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Page 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Page 189: ......