DX-D 600
User Manual
0242F EN 20181218
38
It is customer responsibility to assure that this equipment
and vicinity equipment complies the value of radio
frequency interferences shown in General Regulation for
safety according to IEC 60601-1-2:2007 and 2014 Tables as
described in this section.
The manufacturer is not responsible for any interference
caused by using other than recommended interconnect
cables, accessories and transducers or by unauthorized
changes or modifications to this equipment.
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS
(IEC 60601-1-2:2007 AND IEC 60601-1-2:2014)
The DX-D 600 Radiographic Room is intended for use in the electromagnetic environment specified below.
The customer or the user of the DX-D 600 Radiographic Room should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The DX-D 600 Radiographic Room use RF
energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class A
The DX D 600 Radiographic Room is suitable for
Harmonic emissions
IEC 61000-3-2
Class A
The DX-D 600 Radiographic Room is suitable for
use in all establishments other than domestic and
those directly connected to the public low-voltage
power supply network that supplies buildings
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
power supply network that supplies buildings
used for domestic purposes.
NOTE - In accordance with Standard IEC 60601-1-2:2014, the emissions characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 Class A. If it is used in a residential environment (for which CISPR 11 Class B is normally required) this equipment might not
offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or
re-orientating the equipment.
Summary of Contents for DX-D 600
Page 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Page 2: ......
Page 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Page 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Page 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Page 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Page 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Page 189: ......