DX-D 600
User Manual
0242F EN 20181218
110
THE TABLETOP MUST BE PREVIOUSLY CENTERED AND
WITH THE PATIENT COMPULSORILY LIED DOWN FOR A
SAFE VERTICAL MOVEMENT.
THE RAD TABLE PROVIDES A SAFETY SYSTEM WHICH
STOPS THE VERTICAL MOVEMENT WHEN THE TABLETOP
TRAVEL FINDS AN OBSTACLE.
Before raising or lowering the Tabletop, make sure that no
obstacles are above or below it.
5.3
DR DETECTOR HORIZONTAL MOVEMENT
Press on the DR Detector Brake Control to release the brake and move it manually.
With Automatic Ceiling Suspension, the DR Detector can be moved
automatically when executing any Auto-position or Programmed Position that
has been configured with a DR Detector horizontal displacement.
It is important to have the X-ray Tube exactly centered with the DR Detector. A
density drop at the image edges may appear, indicating an inaccurate alignment.
The Longitudinal center of the Grid and the DR Detector matches with the mark
of the Grid Handle. The X-ray tube can be aligned with the DR Detector by
activating the AUTO-CENTER function on the Ceiling Suspension or moving
manually the X-ray Tube until aligning the Collimator light with the mark of the
Grid Handle.
Depending on the position of the Tabletop, the Grid handle may need to be
pulled out to allow the collimator light to shine on it.
Brake Controls
Fixed Detectors
Removable Detectors
Summary of Contents for DX-D 600
Page 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Page 2: ......
Page 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Page 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Page 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Page 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Page 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Page 189: ......