DX-D 600
User Manual
0242F EN 20181218
99
1.
Select one of the Programmed Positions (PP), use the
INCREASE/DECREASE
Buttons. All parameters values, the
Description and ID Number are displayed.
2.
In case the Programmed Position (PP) involves any movement in the
Longitudinal, Transverse, Alpha and Beta Axes, it must be completed
manually before executing the Auto-position. Press the
AUTO-CENTER
Indicator of the Control Console, a message will be displayed in the
console explaining the corrective action to be completed. It is also
indicated first the Current Position and the Final Position.
Illustration 4-31
Position Message for Semi-automatic System
Final Position
Current Position
Corrective Movement
Direction
3.
Press the
VALIDATE PP
Button to activate the Programmed Position.
4.
Press and hold the
AUTO-CENTER
Button of the Automatic Positioning
Control Box or Remote Control, if available, to get the Ceiling
Suspension on the configured position.
5.
Once the new position is reached the
AUTO-CENTER
Indicator of the
Control Console lights up.
Summary of Contents for DX-D 600
Page 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Page 2: ......
Page 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Page 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Page 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Page 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Page 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Page 189: ......