GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
(IEC 60601-1-2:2007)
This DX-D 600 Radiographic Room is intended for use in an electromagnetic environment specified below.
The customer or Operator of this DX-D 600 Radiographic Room should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2:2007
test level
Compliance level
Electromagnetic environment guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Portable and mobile RF communications equipment
should be used no closer to any part of the DX-D 600
Radiographic Room, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
where ’P’ is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and ’d’ is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
),
should be less than the compliance level in each
frequency range
b
).
Interference may occur in the vicinity of equipment
marked with the following symbol:
d
=
1.2
P
d
=
1.2
P
, 80 MHz to 800 MHz
d
=
2.3
P
, 800 MHz to 2.5 GHz
NOTE 1 -- At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 -- These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the DX-D 600 Radiographic Room is used exceeds the applicable RF compliance level above, this DX-D 600
Radiographic Room should be observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating this DX-D 600 Radiographic Room.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Summary of Contents for DX-D 600
Page 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Page 2: ......
Page 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Page 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Page 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Page 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Page 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Page 189: ......