Remarks for HF-emission and immunity
This device was tested for a normal hospital environment.
WARNING:
The HF-emission and immunity can be influenced by connected
data cables depending on length and the manner of installation.
WARNING:
This device is intended for use by healthcare professionals only.
This device may cause radio interference or may disrupt the
operation of nearby equipment. It may be necessary to take
mitigation measures, such as re-orienting or relocating the
device or shielding the location.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Table 50: Guidance and manufacturer's declaration on electromagnetic
emissions
Emission test
Compli-
ance
Electromagnetic Environment
High frequency RF
emissions in ac-
cordance with
CISPR 11
Group 1
The device uses high frequency energy ex-
clusively for its internal functions. For this
reason, its high frequency RF emission is
very low and it is improbable that neigh-
boring electronic equipment will be disrup-
ted.
High frequency RF
emissions in ac-
cordance with
CISPR 11
Class A
The device is suitable for use in all estab-
lishments other than domestic and those
directly connected to the public low-volt-
age power supply network that supplies
buildings used for domestic purposes.
Harmonic emission
in accordance with
IEC 61000-3-2
Not ap-
plicable
Voltage fluctua-
tions / flickering in
accordance with
IEC 61000-3-3
Not ap-
plicable
The emissions characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) this equipment
192