Automatically switching off the DR 100s
The DR 100s switches off automatically after a specific time since the last
exposure. The time is 30 minutes by default and can be configured from 4 to
60 minutes by an authorized service engineer.
Half a minute before switching off, a sound signal is heard, repeated after 10
seconds and again after 10 seconds. To prevent the system from switching off,
briefly grip the handlebar.
If the battery charge level is too low, the system may be switching off earlier
and without the warning sound signals.
While the system is charging, the system is not automatically switched off, but
the currently logged in user is logged out and the screen goes blank, while the
system remains switched on.
DR 100s | Basic Workflow |
101
0411D EN 20220627 1040