Check if the catch has successfully locked the X-ray tube head, by
trying to push it upward without touching the brake controls.
2.
Lock the device.
• To lock the device using the RFID reader, press and hold the power
button until the green status indicator light starts blinking.
No RFID key must be present on the RFID reader!
1
3
2
1.
Power button
2.
Status indiciator light
3.
RFID reader
• To lock the device using the on/off keypad, press and hold the power
button until the green status indicator light goes out.
2
5
3
1
4
1.
Numeric keypad
2.
Power on/off button
3.
Power status indicator light
4.
Keypad status indicator light
5.
Enter button
Two minutes remain before shutdown. The NX workstation and the X-ray
generator are powered off.
3.
Plug the power cable of the unit into a mains power socket.
4.
Power off the DR detector:
a) turn off the DR detector,
b) remove the battery pack.
5.
Put the DR detector in the storage bin.
The storage bin has locks to secure the DR detector.
96
| DR 100s | Basic Workflow
0411D EN 20220627 1040