Table 41: Maximum settable mAs
focus
small
large
small
large
exposure
time
1 s
3 s
3 s
1 s
minimum
mAs
0.1
0.63
0.1
0.63
kV
maximum
mAs
maximum
mAs
maximum
mAs
maximum
mAs
40
120
250
300
310
50
120
250
300
310
60
120
250
300
310
70
100
250
210
280
80
100
250
210
280
90
80
250
210
200
100
80
250
150
200
110
60
210
150
150
125
60
2100
150
150
Table 42: Monobloc
Model
HF1 R/7 ORB
Maximum power (100 kV - 400 mA)
32 kW
Maximum voltage to the tube
125 kV
Maximum current to the tube
350 mA
Ripple to the maximum power
< 2%
Rise time to the max. power
< 1 ms
Thermal safety
60 °C ± 5° C
Oil volume compensation lung
0.20 dm
3
Maximum thermal capacity
600 kJ
Total thermal safety
900 kJ
Maximum continuous heat dissipation
60 W
Minimum inherent filtration @75 kV
1.4 mm Al
Half-value layer @75 kV
2 mm Al
184
| DR 100s | Product Information
0411D EN 20220627 1040