Automated workflow for warming-up of X-ray tube
The software console provides an automated workflow for warming-up of the
X-ray tube.
1.
Close the collimator blades fully.
2.
Make sure that no one will be exposed.
3.
On the software console, go to the screen with modality controls.
Figure 36: Navigation button for modality controls
4.
Click the button to start the automated workflow for warming-up of the X-
ray tube.
Figure 37: Start the automated workflow for warming-up of the X-ray
tube
A table is displayed with a list of exposures.
1
2
xx
xx
xx
xx
xx
xx
xx
xx
xx
xx
xx
xx
1.
Table with list of exposures
2.
Button to cancel the warming-up procedure
Figure 38: List of exposures for warming-up of the X-ray tube
5.
Perform the exposures and wait for the timer icon to finish between
exposures.
The exposure parameters are set automatically.
DR 100s | Basic Workflow |
83
0411D EN 20220627 1040