112
Portable oxygen analyser analox Mini O2DII
16.6
Tr oubleshooting
SYMPTOM
REASON
SOLUTION
Battery symbol
Flat battery
Ch’ange the battery
Empty screen
Switched off
Bad connection
Switch on
Check screen connection
Check battery connection
Zero reading
Sensor not inserted
Expired sensor
Check connection
Check sensor
Irregular reading
Pressure on the sensor
Radio transmission
Old or faulty sensor
Condensation on sensor
Check flow
Move the machine away
Replace sensor
Dry the face of the sensor
Reading does not change when
the calibration handle is turned
Faulty connections
Sensor broken
Check connections
Replace sensor
Sections of the screen are
missing
Faulty screen
Return to vendor
Impossible to calibrate
Faulty sensor
Sensor not in the air
High altitude
Replace sensor
Check the flow adpter
Calculate percentage equivalent to
20.9
x bar
Varied Reading
Rapid temperature
changes
Do not move the analyser from one
temperature to the next immediately
before use
.
Summary of Contents for MCH 14 TECH NITROX SERIES
Page 31: ...31 Installation 3 Pass the pipe through the hole located in the rear guard Fig 29 30 28 29 ...
Page 80: ...80 Troubleshooting 12 1 1 Low pressure compressor ...
Page 85: ...85 Compression diagram High pressure compressor only air Machine diagrams 14 2 ...
Page 86: ...86 Machine diagrams 14 3 Electrical diagram ...
Page 87: ...87 Machine diagrams ...
Page 104: ...104 Pulley Crankcase and cylinder Low pressure compressor Spare parts ...
Page 105: ...105 Piston rod unit Low pressure compressor Spare parts ...
Page 106: ...106 Cylinder Valve and Filter unit Low pressure compressor Spare parts ...
Page 107: ...107 Pump unit Low pressure compressor Spare parts ...