3110 HELP FILES – SECTION 3
3.3 Calibration for Scripted Variable
Loops
When a waveform segment is part of a scripted
variable loop sequence, the Calibration function
allows a calibration test to be run for each iteration
of the segment in the scripted variable loop.
Note: For instructions on creating a scripted
variable loop sequence, see the topic
“Variable
Loops”
in the section
“Waveforms and Con-
trols.”
To perform the calibration, select a waveform
segment contained within a scripted variable loop.
Then press the Calibration button to open the Cali-
bration Values window. See
Figure 3.7.
The values for all scripted variables in the first
iteration of the segment within the variable loop
will be shown in the Calibration Values window.
Determine the variable that defines the amplitude
of the segment (all scripted variable loops in the
Standards Library use the variable “B” for ampli-
tude levels). Note: Scripted variable loops using
user-created CSV files are not supported by AE
Techron.
Figure 3.7 – Scriped Variable Loop Calibration Values Window
Figure 3.7 – Scriped Variable Loop Calibration Test Window