actiLib Library 4U
– User and Service Guide
Page 82 of 103
B2
Robotic controller communication failed
Reset the unit and retry
operation. After several
occurrences contact
technical support
B3
Robotic controller urgent stop due to a released
magazine
Check if magazine are
completely inserted and
retry operation. After several
occurrences contact
technical support
B4
Cartridge did not transport completely
Gripper could not pick cartridge and CP sensor
not present
After pushing the cart CP sensor still present
B5
Robotic controller doesn’t respond on command
Reset the unit and retry
operation. After several
occurrences contact
technical support
C0
Network initialization failed
Check network cable and
network configuration. If the
error recurs, contact
technical support
C1
Telnet Interface initialization failed
Check network cable and
network configuration. If the
error recurs, contact
technical support
C2
Webserver initialization failed
Check network cable and
network configuration. If the
error recurs, contact
technical support
C6
Ping command did not reached target
Check network cable and
network configuration. If the
error recurs, contact
technical support
C7
Cannot Upgrade from USB
Retry of Firmware upgrade,
if not successful contact
technical support
D0
ROM error.
ROM checksum incorrect
Retry of Firmware upgrade,
if not successful contact
technical support
D1
RAM error.
Power on Self Test ( POST ) has failed,
Retry operation; after
several occurrences contact
technical support
D2
NVRAM error.
R/W operation to NVRAM has failed
Retry operation; after
several occurrences contact
technical support
D3
CTC Error.
Retry operation; after